Global Settings
- Use Ion Source Settings from Tune = False
- Method Duration (min)= 120
- Ion Source Type = NSI
- Spray Voltage: Positive Ion (V) = 2000
- Spray Voltage: Negative Ion (V) = 600
- Sweep Gas (Arb) = 0
- Ion Transfer Tube Temp (°C) = 275
- APPI Lamp = Not in use
- Pressure Mode = Standard
- Default Charge State = 1
Experiment 1
- Start Time (min) = 0
- End Time (min) = 120
- Cycle Time (sec) = 2
Scan MasterScan
- MSn Level = 1
- Use Wide Quad Isolation = True
- Detector Type = Orbitrap
- Orbitrap Resolution = 60K
- Mass Range = Normal
- Scan Range (m/z) = 300-1350
- Maximum Injection Time (ms) = 50
- AGC Target = 2000000
- Microscans = 1
- RF Lens (%) = 30
- Use ETD Internal Calibration = False
- DataType = Profile
- Polarity = Positive
- Source Fragmentation = False
- Scan Description =
Filter MIPS
- MIPS Mode = Peptide
Filter ChargeState
- Include undetermined charge states = False
- Include charge state(s) = 2-6
- Include charge states 25 and higher = False
Filter Dynamic Exclusion
- Exclude after n times = 1
- Exclusion duration (s) = 5
- Mass Tolerance = ppm
- Mass tolerance low = 10
- Mass tolerance high = 10
- Exclude isotopes = True
- Perform dependent scan on single charge state per precursor only = False
Data Dependent Properties
- Data Dependent Mode= Cycle Time
Scan Event 1
- Scan ddMSnScan
- MSn Level = 2
- Isolation Mode = Quadrupole
- Isolation Window = 0.7
- Use Isolation m/z Offset = False
- Multi-notch Isolation = False
- Scan Range Mode = Define m/z range
- Scan Priority= 1
- ActivationType = HCD
- Is Stepped Collision Energy On = False
- Stepped Collision Energy (%) = 5
- Collision Energy (%) = 30
- Is EThcD Active = False
- Detector Type = IonTrap
- Ion Trap Scan Rate = Turbo
- Scan Range (m/z) = 200-1200
- Maximum Injection Time (ms) = 15
- AGC Target = 40000
- Inject ions for all available parallelizable time = False
- Microscans = 1
- Use ETD Internal Calibration = False
- DataType = Centroid
- Polarity = Positive
- Source Fragmentation = False
- Scan Description =